Filter Integrity Tester

FIT-1.1/FIT-1.2

It is microcomputer control of automatic detection, with many kinds of test functions such as bubble point scanning, fast bubble point test and pressure holding method;

More details

Product info

Main Characteristic
Online testing, does not interfere the downstream aseptic conditions
It has system self-check function, with an error report function for the system tightness and wrong operation;
It is microcomputer control of automatic detection, with many kinds of test functions such as bubble point scanning, fast bubble point test and pressure holding method;
Real-time print of test results;
It can be used to test the single core filter and multi-core filter.

Performance Parameter 


Model

FIT-1.1

FIT-1.2

Power supply

170-240V AC, 50/60Hz;100W

170-240V AC, 50/60Hz;85w

Maximum operating pressure

8000mbar

8000 mbar

Minimum inlet pressure

 100mbar

  100 mbar

Dimension

400(W) x 380(D) x 100(H1) x 235(H2)

240(W) x 380(D1) x 280(D2) x 220(H1)

Measuring Range

500-6900mbar

500-6900mbar

Measuring Accuracy

Net Volume Test:± 4%;Bubble Point:± 50mbar ;Diffusion Current:± 4% ;

Sensitivity:± 1.0 mbar ;   Bubble Point:± 75 mbar

Operating condition

Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80%

Ambient Temperature:+5℃ ~ +35℃;Relative Humidity:10-80%

Test time consuming

Net Volume Test:5 min±2min;    Diffusion Current Test:10 min±2min;
Simple bubble point test :15 min±2min;Enhanced Bubble Point Test:20 min±2min ;

Bubble point test :15 min±2min;   bubble point scanning : 20min±2min;Diffusion Current:10min±2min

Printing function

printing, Output test parameters and test results

printing, Output test parameters and test results

History recording function

500 group Test result( Including test curves )

Store 20 groups of test results

Display

size:5.7" TFT; homochromy

size:73*39mm; homochromy

serial connection

serial port :RS232

 

Language

English

English